Model510S
Patient TypeAdult&Child(>3.5kg)
Ventilation ModesA/C-V with sign,A/C-P,SIMV-V,SIMV-P,SPONT/PSV,CPAP,MANUAL
EnhancementsApnea back-up ventilation,Manual breath,Alarm silence
Parameter Setting
Tidal Volume(Vt)

0-2000 mL

Pinsp5-50 cmH₂O
Psupp0-50 cmH₂O
CPAP(NIV)0-30 cmH₂O
PEEP0-30 cmH₂O
Pressure trigger-20-0 cmH₂O
Flow trigger2-30 L/min
Frequency(f)1-120 bpm(A/C),1-40 bpm(SIMV)
I:E ratio4:1-1:10
Pause time(Tpause)0-5 s
O₂ concentration40-100%
Inspiration flowMax. 90 L/min
Monitoring DataMV,VT,Frequency(f),FiO₂,Ppeak,PEEP
Waveform DisplayP-T,F-T
Alarm Setting
High MVOFF,1-25L
Low MVOFF,0-24L
High FiO₂OFF,50-100%
Low FiO₂OFF,35-99%
High Paw6-80 cmH₂O
Low Paw0-40 cmH₂O
AC power failureGas supply down
O₂ deficiencyMV high
Low batteryNo VT
ApneaCP high
LanguageChinese,English,Spanish,Italian,Turkish,Other languages
Environmental Requirement
Temperature
Operating condition-18-50 ℃
Storage condition-20-60 ℃
Relative humidity
Operating condition
15-95%
Storage condition10-95%
Barometric pressure
Operating condition70-110 kPa
Storage condition50-110 kPa
Electrical
AC voltage100-240V,50/60Hz
DC voltage12V
Battery run timeMinimum 4.5 hours
Battery typeNickel-cadmium battery
Gas Supply
Supply pressure(O2)250-600 kPa
FlowMax. 180 L/min
Physical Specifications
Main unit weight
3 kg(exclude trolley)
Dimension(H*W*D)168*156*300 mm
LCD screen5'' TFT
Communication PortsRS232


Stable Condition Phase 

• PRVC and BIVENT employ lung-protective strategies, delivering intelligent ventilation 

• Comprehensive lung mechanics monitoring include compliance, airway resistance, PEEPi and time constant 

• Three waveforms & three loops with user-friendly display provide a continuous monitoring of the patient's condition 


Weaning Phase 

• Various ventilation modes enhance the weaning process 

• The unique trigger and leakage compensation system safeguards each and every  patient breath resulting in smooth and comfortable breathing, avoiding extra workload on the patient and promoting recovery 

• RSBI and WOB provide accurate reference for weaning 


Complete Ventilation Modes 

1.The device hosts six ventilation modes.  

2.Besides the conventional Volume-Controlled ventilation  mode, it also has Pressure-Controlled ventilation mode,  

3.SIMV and CPAP modes.  The machine can provide a precise tidal volume.  

4.It also provides inspiration hold and manual breath,  one-button operation, and is thus incredibly convenient. 


Ideal O2 Delivery 

1.With its advanced technology, the range of FiO2 can be  set from 40% to 100%.  

2.Furthermore, this range can be adjusted continuously.